The Journal of Internal Korean Medicine 2013;34(3): 231-239. |
맥진기 판독의 신뢰도 및 파형의 재현성 연구 |
박승찬, 이지혜, 이혜윤, 조민경, 김도형, 김소연, 최준용, 한창우, 박성하, 홍진우, 이인, 권정남 |
부산대학교 한방병원 한방내과 |
Study on Reliability of Interpretation and Reproducibility of a Pulse Analyser |
Seung-chan Park, Ji-hye Lee, Hye-yoon Lee, Min-kyoung Cho, Do-hyung Kim, So-yeon Kim, Jun-yong Choi, Chang-woo Han, Seong-ha Park, Jin-woo Hong, In Lee, Jung-nam Kwon |
Dept. of Korean Internal Medicine, Pu-San National University Korean Medicine Hospital |
|
Published online: September 30, 2013. |
|
|
ABSTRACT |
Objectives: This study was performed to evaluate inter-rater and intra-rater reliability of interpretation and reproducibility of a pulse analyser (MAXMAC27-Plus).
Methods : 38 of 40 volunteers completed the pulse analysis consecutively. Three Korean medical doctors who had at least 2 years of clinical experience interpreted the pulse waves for 3 aspects of size, depth and shape, then inter-rater reliability and crude agreement was obtained. Reinterpretation was done 2 weeks later and intra-rater reliability and crude agreement was obtained. Intra-rater reliability and crude agreement between 1st and 2nd measurement was calculated. Cohen's weighted kappa for size, Cohen's kappa for depth and shape were used as statistical analysis.
Results: Inter-rater reliability of size, depth and shape among 3 raters was 0.598, 0.604, and 0.312, respectively, showing moderate to substantial agreement. Average intra-rater reliability between 1st and 2nd interpretation of size, depth and shape was 0.806, 0.705, and 0.638, respectively, showing substantial to almost perfect agreement. However, intra-rater reliability between consecutive measurements of size, depth and shape was 0.221, 0.121, and 0.194, respectively, which showed only poor to fair agreement.
Conclusions: Intra-rater and inter-rater reliability of one pulse wave showed relatively high concordance. Training by a clinical expert may effect better concordance among raters. Test-retest reliability showed poor agreement. Improvement of measurement technique and device performance will be needed. |
Key words:
pulse analyser, inter-rater reliability, intra-rater reliability, reproducibility |
|
|
|